Standard features of the X1000 X-Ray Inspection System include; A 2”x2” digital flat panel x-ray detector, 80 kV- 7 micron x-ray tube, Z axis movement of the x-ray tube and image detector, computer controlled kV and mA settings, computer controlled variable speed X-Y stage and full featured image processing CPU with 17 inch LCD flat panel monitor. The standard 1000 does not have a tilt or oblique viewing feature; there is an optional 360 degree rotation sample stage that
1-1. Safety Chamber |
Dimensions - L*W*H : 43” * 29.5” * 55” |
Weight : Approx 1000 lbs |
Power requirements : 50/60Hz,200VA,100 ~ 240VAC +/- 10% |
X-ray leakage : less than 0.5mR/hour ( 100% protection safety chamber ) |
|
1-2. Sealed Type End window Micro Focus 80kV X-Ray Tube |
Integrated 80kV Micro Focus X-Ray Source |
Operating Voltage Range : 80 kV |
Maximum Beam Current : .12 mA |
Focal Spot Size : ≤8 ㎛ |
Tube Life Time : Approximately 5 Years |
1-3. Flat Panel Digital Detector |
High performance 12 bit |
Resolution 1024 x 1024 |
Speed 1.2 fps |
Video Displayed on 19” LCD monitor |
Total system magnification 125X, Geometrical magnification 15X |
1-4. Image Processor Capture X on the PC based Windows XP workstation |
A software package designed specifically for real-time x-ray inspection systems |
creates a high-performance PC-based image workstation with Windows XP |
Noise reduction by frame averaging up to 256 most recent frames |
CPU Speed : P4-1GHZ or higher |
Memory : 256MB/133MHz SDRAM(min) or higher |
Graphics card : 128MB AGP(min) |
System : Windows XP® |
1-5. 3 Axis Manipulator / Sample Stage |
Special designed computer controlled manipulator |
X,Y Axis stroke : 7.8” x 7.8” |
Z Axis stroke: 8.6” Sample Stage: 12.5” x 9.8”
|
1-7. Installation |
WEwill provide installation and calibration service at the customers location |
Installation includes assistance in registering your new x-ray system with local and state agencies where applicable. |
One (1) on-site radiation survey at the time of installation with supporting documentation |
1-8. Training |
One day (6 hours maximum) of training will be provided by Scienscope, training will include; |
Basic Radiation Safety |
X-Ray system control functions |
X-Ray image processing software training |
Basic X-ray signature analysis training |
Hands-on sample analysis using your typical samples |
Training certificates for all attendees
1-8. Two Hours On Line Training (First 12 Months) we will provide two hours of online training during the first 12 months. Online training can be used at the customer’s discretion and divided up into 15 minute, 30 minute and one hour increments. Additional online training can be purchased at a minimal cost.
|
1-9. Warranty |
Weprovides a twelve month (12) warranty or 2000 hours (whichever comes first), parts and labor |